SL 243

True Double Beam Atomic Absorption Spectrophotometer

 

Salient Features & Benefits
  • 7” High Resolution Touch Screen LCD Color Graphical Display SUPPORTS USB PEN DRIVE TO STORE / RETRIEVE MASS SPECTRAL DATA .
  • BUILT-IN PC Control with Advanced capability for Data Processing, Processing, Storage, Retrieval, Interpretation, etc.
  • Background correction with D2 lamp source.
  • Auto Burner Height Optimization, and maximum ABS adjustment of burner through membrane keyboard .
  • Auto Flame control.
  • Integral Microcontroller based auto gas box with all safety features.
  • Supports USB Printing with PCL Class Printers.



  • Applications
    For analysis of various metallic constituents in
  • Soil
  • Fertilizer
  • Micro Nutrients
  • Plants
  • Pesticide
  • Foods
  • Water
  • Biological
  • Matter
  • Metallurgy
  • Plating Coal
  • Petrochemicals
  • Glass
  • Cement
  • Steel
  • Others


  • Industries using SL-243 True Double Beam Atomic Absorption Spectrophotometer
    Agriculture Government Food & Beverages
    Clinical Environmental Chemical & Pharma



    Displaying the SL-243 True Double Beam Atomic Absorption Spectrophotometer

    Specifications

    SPECTRAL
    Range 185 to 900 nm
    Bandwidth Auto select in 8 steps from 0.1 to 2.5 nm
    PHOTOMETRIC
    Range 0 to 2 Abs
    LIGHT SOURCE Hollow cathode lamp (HCL)
    BACKGROUND CORRECTION Deuterium Lamp (D2 ) - Optional
    BURNER Titanium construction, Premix, laminar flow, 100 mm burner for Air  – Acetylene (C2 H2 )
    FUEL Acetylene (C2 H2 )
    OXIDANT Air/Nitrous Oxide (N2O)
    FLOW CONTROL Optional: Integral auto gas control system and panel flow meters, Automatic flame change over.
    SAFETY FEATURES Full safety interlock including pressure sensors, power failure, Protection, auto ignition, flame failure detection & burner recognition.
    NEBULIZER Pneumatic, with 90% platinum and 10% iridium alloy capillary
    Sample feed rate 3 to 8 ml/min (adjustable)
    DATA PROCESSING
    Built-In Advanced Capability 1. Absorbance, 2. Concentration (standard addition and calibration), 3. Curve fit (Quadratic and Least Squares), 4. Energy Profile, 5. Emission, 6. Printing MS Office Tools
    MONOCHROMATOR Czerney Turner 1/3 m with 1800 lines/mm holographic diffraction grating and hard coated optics with full temperature compensation. Reciprocal, linear, dispersion is 1.5 to 2.0 nm/mm
    DETECTOR Wide range Photo Multiplier Tube (PMT)
    SYSTEM UPGRADATION Instrument can be upgraded to hydride and cold vapor generator
     